No. The turbine doesn’t need oil, but the handpiece still requires daily external cleaning, internal line clearing with isopropyl alcohol and syringe air before each sterilization cycle, and weekly push-button chuck cleaning using a product like Star’s Junk-Out kit. Skipping those steps leads to chuck debris buildup and seal degradation.